BFMUV-2000 Microspectrophotometer
INSTRUMENT CHARACTERISTICS
·Intelligent Android operating system, 7 inch capacitive touchscreen, multi-touch, special APP software, more intuitive interface, user-friendly design.
·Cuvetteslot is more convenient for the detection of bacteria/microbes and other culture liquid concentration.
·Only 0.5 ~ 2μL sample is required for each test.After the test, you can also use APP software, with a more intuitive interface and user-friendly design.
·The sample is directly added to the sample testing platform without dilution. The test can be completed in 8s and the results can be directly output as
the sample concentration.
·Xenon flash lamp, 10 times life (up to 10 years). Boot without preheating, direct use, can be detected at anytime.
·The sample is directly put on the sampling platform, without dilution, the sample concentration can be measured for the conventional UV-visible spectrophotometer 50 times, the results directly output as the sample concentration, without additional calculation.
·Stable and fast USB data output, easy to export data for corresponding analysis.
·The instrument does not need computer online, single machine to complete sample testing and data storage.
·Image and table storage format, table compatible with Excel, convenient for subsequent data processing, support JPG image export.
·Driven by high-precision linear motor, the precision of optical path can reach 0.001mm, and the absorbance test has high repeatability.
ERFORMANCE PARAMETER
Name | Microspectrophotometer |
Model | BFMUV-2000 |
Wavelength range | 200 ~ 800nm; Colorimetric mode (OD600 measurement) : 600±8nm |
Sample volume | 0.5~2.0μl |
Optical path | 0.2mm (high concentration measurement); 1.0mm (ordinary concentration measurement) |
Light source | Xenon flash lamp |
Detector | 2048 units linear CCD display |
Wavelength accuracy | 1nm |
Wavelength resolution | ≤3nm(FWHM at Hg 546nm) |
Absorbance accuracy | 0.003Abs |
Absorbance | 1%(7.332Abs at 260nm) |
Absorbance range (equivalent to 10mm) | 0.02-100A; Colorimetric mode (OD600 measurement) : 0~4A |
Test time | <8S |
Nucleic acid detection range | 2~5000ng/μl(dsDNA) |
Data output mode | USB |
Sample base material | quartz fiber and high hard aluminum |
Power adapter | 12V 4A |
Power consumption | 48W |
Power consumption during standby | 5W |
Software operating system | Android |
Size(mm) | 270×210×196 |
Weight | 3.5kg |